The HI96822 is a rugged, portable digital refractometer
designed for the measurement of salinity of natural or artificial seawater. The
HI96822 displays results in three popular measurement units: Practical Salinity
Units (PSU), salinity in parts per thousands (ppt) and specific gravity (S.G.
(20/20)). The instrument's high accuracy and simple operation gives reliable
results each and every time. All readings are automatically compensated for
temperature variations and displayed within a 1.5 second response time. The
sealed flint glass prism and stainless steel well are easy to clean. Just wipe
with a soft cloth in preparation for the next sample.
- Sample
size as small as 2 metric drops (100 μl)
- Sealed
stainless steel well with high-grade optical prism made of flint glass
- Fast
1.5 second response time for temperature compensated readings
The HI96822 Digital Refractometer eliminates the uncertainty
associated with mechanical refractometers by measuring the refractive index to
determine the salinity of natural and artificial seawater including brackish,
ocean water or saltwater aquarium. The instrument converts the refractive index
of a sample to PSU, ppt and specific gravity. These conversions are based on
respected scientific publications using the physical properties of seawater
(not sodium chloride). The HI96822 is an easy to use tool for measuring
salinity in natural and artificial seawater whether in the field, lab or at
home.
- Designed
for seawater salinity analysis (not NaCl)
- Automatic
Temperature Compensation
- High
accuracy measurements displayed as PSU, ppt and specific gravity
Making a Standard Solution
Sodium Chloride solutions can be used to check the accuracy
of the meter. The table below lists two Sodium Chloride solutions and their
expected ppt Seawater value.
To make a Standard NaCl Solution (g/100 g), follow the procedure below
- Place
container (such as a glass vial or dropper bottle that has a cover) on an
analytical balance.
- Tare
the balance.
- To
make an X NaCl solution weigh out X grams of high purity dried Sodium
Chloride (CAS #: 7647-14-5: MW 58.44) directly into the container.
- Add
distilled or deionized water to the beaker so the total weight of the
solution is 100g.
Example Standard NaCl Solution
NaCl (g) Water (g)
Total Expected Seawater value (ppt) 3.5% NaCl 3.50 96.50 100.000 34 10%
NaCl 10.00 90.00 100.000 96
|