The Milwaukee MC120 PRO High Range pH monitor offers lab
grade accuracy with an easily viewable continuous check on pH levels with
alarms.
Design features
- Fast
easy to read results with large digital LCD readout.
- Accuracy
to ± 0.2 pH with a full 0.0 pH to 14.0 pH range.
- Set
point range of 5.5 to 9.5 pH with alarm activated when reading is higher
or lower than set point
- Comes
with full mounting kit including probe holders.
- Lab
grade probe included with a double junction, gel filled design for low
maintenance.
Use and Care
pH meters usually start performing poorly because of
problems with the probe. The two parts of the pH probe that cause problems are
the glass sensing bulb and the reference junction.
Sensing bulb
The glass bulb loses sensitivity with use and will
eventually fail. This is true of all pH probes.
Reference junction
The reference junction is a small whole that allows the
meter to compare the sample to a reference. The reading is generated based on
the electrical difference between the sample and reference. If the junction
hole gets clogged, the pH electrode will no longer function.
You can extend the life of your pH electrode in the
following ways.
Storage
The pH probe needs to be kept hydrated. Long periods of dry
storage will damage the sensitivity of the probe. Allowing the probe to dry out
may also result in the junction hole getting clogged.
Storing in pH probe storage solution or pH calibration
buffers will help address both these issues.
Do not store in tap water and DI water. This will damage the
sensitivity of the probe.
The best way to store the probe is with the probe's cap
filled with storage solution or calibration buffer and the cap tightened to
prevent leakage. It is also recommended that to store the electrode upright to
further reduce the potential for leakage.
Cleaning
Residue from the sample can impact the sensitivity of the
sensing bulb and clog the junction. This is especially true if the sample has a
lot of organic material. We recommend regularly soaking the probe in cleaning
solution.
Calibration
The pH sensing bulbs become less sensitive over time so make
sure that you are calibrating regularly to keep your PRO pH meter accurate.
Specification
- pH
Range: 0.0 - 14.0 pH
- pH
Resolution: 0.1 pH
- pH
Accuracy: ± 0.2 pH
- Calibration:
Manual 2 points
- Setpoint
range: 5.5 to 9.5 pH
- Alarm
Active: when reading is higher or lower than set point
- pH
Electrode: MA911B/2 (included)
- Environment:
0 to 50°C / 32 to 122°F; max RH 95%
- Power
Supply: 12 VDC adapter (included)
- Mounting
kit (Included)
- Dimensions:
5.9 x 3.2 x 1.35 inches
- Weight:
6.3 ounces (meter only)
|